Flexural Testing Acrylic Resin Cements Used in Orthopaedics (ISO 16402)

During the implantation process, bone cements are used to fill the space between the implant and the bone, and their long-term mechanical stability is essential for success of the implant. By undertaking static and cyclic loading, the mechanical performance of these bone cements can be established.

ISO 16402 ‘Implants for surgery – Acrylic resin cement - Flexural fatigue testing of acrylic resin cements used in orthopaedics’ provides a guideline for flexural testing of these cements in a laboratory situation. It consists of a method for both quasi-static bend testing and flexural fatigue.

A miniature 4-point bend fixture is used within into a fluid bath filled with Ringer’s solution and maintained at 37° C. After a quasi-static bend test is performed, fatigue testing is carried out to generate a Stress level/Number of cycles (S/N) curve.

We recommend the use of either an ElectroPuls™ All-Electric Dynamic Test Instrument or 8870 servohydraulic fatigue test system for undertaking both the quasi-static and the cyclic loading of these bone cements.

We recommend that you review the standard to fully understand its requirements.

Instron has a variety of solutions for testing of orthopaedic implants and devices or biomaterials. Find out more.


このソリューションについて

関連試験規格: ISO 16402
試験片タイプ: 矩形
Materials: バイオメディカル | バイオマテリアル
試験形式: 曲げ | 高サイクル疲労
ビジネスセクター: 学術研究 | バイオメディカル/生体材料/医療
4 Point Submersible Bend Fixture
You are viewing a static image in place of an informative flash animation. Please download Adobe® Flash Player and refresh this page to view the animation.
このソリューションへのご質問などは・・・
お問合せください

関連資料

5900 シリーズ試験結果のゆるぎなき信頼性   5900 シリーズ試験結果のゆるぎなき信頼性
ElectroPuls™ 試験システム   ElectroPuls™ 試験システム
Solutions for Biomedical Testing   Solutions for Biomedical Testing
整形外科分野および骨接合術用器具   整形外科分野および骨接合術用器具

弊社デモルーム

インストロンジャパンのデモルームで、実際に試験機をご覧頂くことができます。また、試験片をお持ちいただき、適切なグリップなどのアクセサリーについてご相談いただくことも可能です。
詳しくはお問合せください。