Flexural Strength Testing of Dental Ceramics (ISO 6872)

Ceramics materials are used in modern dentistry in a variety of different applications including fillings, crowns, veneers, implants, and dental brackets.

ISO 6872 "Dentistry - Ceramic Materials" specifies the requirements for ceramic materials used in dentistry and also defines the basic methods for flexural strength testing. The standard specifies that either three or four-point bending is permissible.

Testing can be carried out on either a Universal Electromechanical Testing Machine or an ElectroPuls™ All-Electric Dynamic Test System, using Bluehill® Software to set up, run, and analyze tests.

To accommodate the small specimen size, we used a miniature bend fixture with the ability to incorporate different anvil diameters. The test can be undertaken in air or can be submersed into a fluid bath for simulation of in vivo conditions.

Our bend fixture provides specialized features for testing according to ISO 6872. The fixture is engineered to ensure high precision of span distances and centering, and addresses the issues of alignment and parallelism that are vital for these tests. The unit also features and includes various sized anvils with mounting in a V-slot for high anvil parallelism.

We recommend that you review the standard to fully understand its requirements.

Interested in other mechanical tests for dental materials or dental implants? We have a variety of different testing equipment solutions for dentistry.


このソリューションについて

関連試験規格: ISO 6872
試験片タイプ: 矩形 | Beam
Materials: バイオメディカル | バイオマテリアル
試験形式: 曲げ | 高サイクル疲労
ビジネスセクター: 学術研究 | バイオメディカル/生体材料/医療 | 受託試験サービス | 官庁/防衛関連
Flexural Strength Testing of Dental Materials
You are viewing a static image in place of an informative flash animation. Please download Adobe® Flash Player and refresh this page to view the animation.
このソリューションへのご質問などは・・・
お問合せください

関連資料

歯科材料および歯科インプラント   歯科材料および歯科インプラント

弊社デモルーム

インストロンジャパンのデモルームで、実際に試験機をご覧頂くことができます。また、試験片をお持ちいただき、適切なグリップなどのアクセサリーについてご相談いただくことも可能です。
詳しくはお問合せください。