Test method for liquid leakage at syringe piston, ISO 7886-1 Annex D

The standard ISO 7886-1 : 1993 Annex D specifies a test method for liquid leakage at syringe piston under compression. The procedure includes the following steps: 1. Draw into the syringe a volume of water exceeding the nominal capacity of the syringe. 2. Expel air and adjust the volume of water in the syringe to the nominal capacity. 3. Seal the syringe nozzle using the cap provided. 4. Apply and axial force and maintain it for 30 seconds. 5. Examine the syringe for leakage of water past the piston seals.

For this particular test, a force value 20% below the level when a failure of the cap was registered was the force selected to constantly apply during 30 seconds. This period started after the selected force level was reached for the first time. After this test, a leakage at the nozzle cap should be visually detected, a leakage either partial or total can be also monitored if a drop in the force level is noticed during the test. The Instron Syringe fixture CP102517 provides an ideal solution for a liquid leakage test, the Instron® software Bluehill® also provides the required control tools to easily perform a force hold during this test.


このソリューションについて

関連試験規格: ISO 7886-1
試験片タイプ: 注射器
Materials: バイオメディカル
試験形式: 圧縮
ビジネスセクター: バイオメディカル/生体材料/医療
You are viewing a static image in place of an informative flash animation. Please download Adobe® Flash Player and refresh this page to view the animation.
このソリューションへのご質問などは・・・
お問合せください

関連資料

Bluehill 3 試験ソフトウエア   Bluehill 3 試験ソフトウエア
Syringe Test Fixture   Syringe Test Fixture
Syringe Test Fixture for ISO-7886-1   Syringe Test Fixture for ISO-7886-1

弊社デモルーム

インストロンジャパンのデモルームで、実際に試験機をご覧頂くことができます。また、試験片をお持ちいただき、適切なグリップなどのアクセサリーについてご相談いただくことも可能です。
詳しくはお問合せください。